Meditech Patient Monitor Model MD9012 Monitor 105 dB

Zirar Enterprises

No Price Available

Product Details:

Manufacturer: Meditech
Condition: new
Purpose: Sale
Model number: MD9012
Location: Lahore

Description:

Meditech Patient Monitor

PATIENT MONITOR MD9012 MEDITECH

Patient Monitor MD9012 Meditech is 12 inch TFT display high resolution. Automatic temperature drift treatment of NIBP measuring patented technology; Friendly and configurable user interface; Powerful data review Maximum 720-hour graphic and tabular trends, up to10000 pieces of NIBP record storage, Wired or wireless networkable with Central Monitoring System; On-line upgrading keeps you updated.


FEATURES:

  • High resolution 12 Inch color TFT display
  • 12 Channel waves.9 Parameters: ECG, Resp, NIBP, SpO2, TEMP,
  • Light weight, compact and portable
  • Built-in rechargeable lithium battery
  • Unique human voice alarm
  • Rich screen layouts: Standard, Big font, Trend table, NIBP trend, ECG full lead, OxyCRG, IoC Layout
  • Powerful storage: Maximum 720 Hour graphic and tabular trends of all
  • parameters, up to 10000 NIBP review, 200 alarm review ,2 Hours of all wave forms
  • System software can be upgraded through network
  • Data and waveforms color be adjustable
  • Arrhythmia analysis and S-T segment analysis
  • Suitable for adult, pediatric and neonatal patient


Options:

  • Invasive Blood Pressure (IBP)
  • Carbon dioxide (CO2) Module
  • Thermal Printer
  • Touch Screen


SPECIFICATION:

ECG

  • Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)
  • Lead selection I, II, III, avR, avL, avF, V,
  • Waveform 2 ch
  • Lead mode 3 Leads ( R, L, F or RA, LA, LL)
  • Lead selection I, II, III,
  • Waveform 1 ch
  • Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
  • HR and Alarm
  • Range
  • Adult 15 ~ 300 bpm
  • Neo/Ped 15 ~ 350 bpm
  • Accuracy   1 Percent or   1bpm,which great
  • Resolution 1 bpm
  • Sensitivity   200 (uV )
  • Differential Input Impedance   5 M 
  • CMRR
  • Monitor   105 dB
  • Operation   105 dB
  • Diagnosis   85 dB
  • Electrode offset potential  300mV
  • Leakage Current   10 uA
  • Baseline Recovery   3 S After Defi.
  • ECG Signal Range  8 m V ( Vp-p )
  • Bandwidth
  • Surgery 1 ~ 15 Hz
  • Monitor 0.5 ~ 35 Hz
  • Diagnostic 0.05 ~ 100 Hz
  • Calibration Signal 1 (mV ), Accuracy ? 5 Percent
  • ST Segment Monitoring Range
  • Measure and Alarm -2.0 ~ +2.0 mV
  • ARR Detecting
  • Type ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT 2, PVC, TACHY, BRADY,MISSED BEATS, PNP, PNC
  • Alarm Available
  • Review Available


RESPIRATION

  • Method Impedance between R-F(RA-LL)
  • Differential Input Impedance  2.5 M 
  • Measuring Impedance Range: 0.3~5.0 
  • Base line Impedance Range: 0 – 2.5 K 
  • Bandwidth 0.3 ~ 2.5 Hz
  • Resp. Rate
  • Measuring and Alarm Range
  • Adult 0 ~ 120 rpm
  • Neo/Ped 0 ~ 150 rpm
  • Resolution 1 rpm
  • Accuracy  2 rpm
  • Apean Alarm 10 ~ 40 S


NIBP

  • Method Oscillometric
  • Mode Manual, Auto, STAT
  • Measuring Interval in AUTO Mode 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240,480 (Min)
  • Measuring Period in STAT Mode 5 Min
  • Pulse Rate Range 40 ~ 240 bpm
  • Alarm Type SYS, DIA, MEAN
  • Measuring and alarm range
  • Adult Mode
  • SYS 40 ~ 270 mmHg
  • DIA 10 ~ 215 mmHg
  • MEAN 20 ~ 235 mmHg
  • Pediatric Mode
  • SYS 40 ~ 200 mmHg
  • DIA 10 ~ 150 mmHg
  • MEAN 20 ~ 165 mmHg
  • Neonatal Mode
  • SYS 40 ~ 135 mmHg
  • DIA 10 ~ 100 mmHg
  • MEAN 20 ~ 110 mmHg
  • Resolution
  • Pressure 1mmHg
  • Accuracy
  • Pressure
  • Maximum Mean error  5mmHg
  • Maximum Standard deviation  8mmHg
  • Overpressure Protection
  • Adult Mode 297 3 mmHg
  • Pediatric Mode 240 3 mmHg
  • Neonatal Mode 147 3 mmHg


SpO2

  • Measuring Range 0 ~ 100  Percent
  • Alarm Range 0 ~ 100  Percent
  • Resolution 1  Percent
  • Accuracy 70 Percent ~ 100 Percent  2  Percent
  • 0 Percent ~ 69 Percent unspecified
  • Actualization interval about 1 Sec.
  • Alarm Delay 10 Sec.
  • Pulse Rate
  • Measuring and Alarm Range 0~254bpm
  • Resolution 1bpm
  • Accuracy  2bpm

TEMPERATURE

  • Channel 1
  • Measuring and Alarm Range 0 ~ 50 C
  • Resolution 0.1~C
  • Accuracy  0.1~C
  • Actualization interval about 1 Sec.
  • Average Time Constant   10 Sec.


IBP

  • Label ART, PA, CVP, RAP, LAP, ICP, P1, P2
  • Measuring and alarm range
  • ART 0 ~ 300 mmHg
  • PA -6 ~ 120 mmHg
  • CVP/RAP/LAP/ICP -10 ~ 40 mmHg
  • P1/P2 -10 ~ 300 mmHg
  • Press Sensor
  • Sensitivity 5 uV/V/mmHg
  • Impedance 300-3000 
  • Resolution 1 mmHg
  • Accuracy  2 Percent or  1mmHg, which great about 1 Sec.